SERVICING & CLEANING RECOMMENDATIONS
- Terminals:
The terminals can be cleaned with a cloth soaked
in disinfectant. - Adhesive electrodes:
Rehydrate before and after use.
If necessary, use disinfectant.
Leave on the original support base, in a plastic
bag at between 5°C and 10°C.
CONDITIONS FOR USE
- Temperature: 5°C - 40°C (max. 35°C while charging)
- Relative humidity: 20% - 65%
STORAGE CONDITIONS
- Storage temperature: 0°C - 45°C
- Relative humidity in the storage area: 10% - 90%
- Keep in a dry location, away from direct sunlight.
RECOMMENDATIONS :
Make sure you comply with national regulations when you dispose of your product at end-of-life. Please dispose of this product at a location designed for this purpose in order to ensure the protection of the environment.
CLASS II MAINS POWER
- Reference : MN0151-E050300
- AC 100-240V - 50/60Hz - 0,5A
DC 5V
3A
Only use the mains power unit supplied with the device. Please replace this unit should it fail (contact
SPORT-ELEC®, or order at www.sport-elec.com).
ELECTRODE DATA
Class I medical device (directive 93/42/EEC) - Adhesive, hydrophilic and hypoallergenic electrodes
Reference .. . . . . . . . . . . . . . . . . EASF110X71 - EAEXP100X50
Contact surface .. . . . . . 67,5x47mm (x2) - 100x50mm (x2)
Estimated life cycle .. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40 usages
Do not hesitate to order FREE ACTION adhesive
electrodes at www.sport-elec.com..
TERMINAL DATA
TCurrent type .. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . two-phase
Number of programs . . . . . . . . . 14 (8 “SPORT” + 6 “HEALTH”)
Number of program variations (or phases).. . . . . . . . . . . . . 255
Number of adjustable independent channels: .. . . up to 4Maximum intensity per channel up to 1kΩ.. . 90mA
Frequency range .. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0.5-120 Hz
Pulse range .. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 70-360 μs
Power .. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Rechargeable LiPo battery
- 3.7V - 1020 mAh (remote control)
- 3.7V - 420 mAh (motors)
Remote control dimensions .. . . . . . . . . . . . 79 x 135 x 24 mm
Remote control weight.. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 146 g
Motor dimensions .. . . . . . . . . . . . . . . . . . . . . . . . . . . 78 x 52 x 20 mm
Motor weight.. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51 g
Contact sensor .. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Yes
- (channel power gauge > white)
Automatic stop at the end of each session.. . . . . . . . . . . . . Yes
Start safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Yes: AutotestClassification.. . . BF type: device in contact with the patient, excluding the heart muscles.
RoHS..Devices are constructed without materials with a potential harmful impact on the environment.
EC declaration of conformity.